Automatic Steam Sterlizier Fully Automatic Vertical GMP

Model No:JRIC –39B

 
     
  » Specification  
     
 

Application: For use in Microbiology, QC & Research laboratory - Pharma industry

 
 

Double wall design has single chamber for steam and water

Fully automatic operation; cycle begins by press of the START button
Initial air purging cycle; at the start air is automatically removed from the chamber
Microprocessor based digital temperature indicator controller controls temperature / pressure precisely at set value
Temperature sensor: PT – 100
In built digital timer; timer can be adjusted as per sterilization load requirement
Automatic steam exhaust at the end of cycle
Low water level heater safeguard.
Safety high pressure release valve
Safety high temperature cut off. Included if ordered separately
Working chamber made of stainless steel LM 304 grade of 2mm thickness
Outer cover made of stainless steel 1mm thick
Lid, flange & bottom sheet also made of stainless steel. All joints argon welded
Joint less silicon gasket. Heavy duty industrial flange heater
Pressure range: 15 to 22 PSI, factory set at 15 PSI
Temperature range: 121°C to 125°C
Temperature resolution: 0.1°C / temperature accuracy: ±0.5°C
Lid fitting pressure gauge 0-30 PSI, safety spring loaded pressure valve, steam release valve
Hydraulically tested at 60PSI
Foot lifting arrangement included for size 18”x 24”, 18” x 30” & 22” x 30”
Electrical: 230V / 15A / 50 Hz
Calibration certificate by ERTL – Mumbai with traceability to NPL – New Delhi
Provided with  DQ, IQ, OQ,  documentation
 
 
We understand customer needs as well as aim at exceeding general expectation
 
Constantly strive to strengthen and sustain quality levels
 
36 years old organization with sound integrity and commitment
 
  >> Read more
Automatic Steam Sterlizier
 
     
     
 

Size of Working Chamber

Capacity
Rating
External Dimmentions
in Cms.
Shpg Wt in kgs

in inches
Dia x Height

in cms
Dia x Height

Liters
KW
   
12 x 20 30 x 50 35 2.0 75 x 48 x 105 60
14 x 22 35 x 55 52 3.0 48 x 80 x 120 80
18 x 24 45 x 60 95 4.0 70 x 90 x 135 110
18 x 30 45 x 75 125 4.0 70 x 90 x 145 135
22 x 30 55 x 75 175 6.0 85 x 110 x 145 140
 
     
 
Optional  Extras / Accessories
a) Digital high temp. Safety controller cum indicator which can be set 2°C above the set temperature for safety purposes and will cut-off the heater with audio-visual alarm ensuring temperature within chamber doesn’t exceed safety threshold.
b) Printer interface facility to record data of 1 temperature and 1 pressure location within the chamber, can print date, time, temperature when connected to EPSON LX 300 Dot matrix printer. (Not supplied with printer)

c) Printer interface facility to record data of 3 temperature and 1 pressure location within the chamber, can print date, time, temperature when connected to EPSON LX 300 Dot matrix printer. (Not supplied with printer)

d) PC interface with software
e) 24 column printer (Imported) to be connected to above printer interface
f)– do – with FO indication with standard PC software. (Not supplied with printer)
g) Basket made of thick SS wire mesh rods
 
     
  Temperature Mapping/Validation  
     
 

Autoclave : Performance qualification with temperature mapping on site.

Temperature mapping consist of totally six cycles. Three cycles to demonstrate the uniformity of temperature within empty chamber, and the other three cycles in loaded chamber condition. This is done to demonstrate that temperature distribution in the loaded chamber is acceptable and that the items within the chamber receive a uniform heat treatment.

Data is collected in printed format from up to six different points within the chamber.
Temperature uniformity is considered acceptable if the deviation is ±1ºC of set sterilization temperature.

 

Note: Other alternative methods of validation can be used such as Biological indicator studies. These can be done by the laboratory staff.

Note: Every OSWORLD Autoclave is validated prior to dispatch at the factory. Those customers who do not require on site mapping may request for FREE copy of factory mapping cycle data sheet.