a) PLC based system : PLC with HMI 3.7” touch screen grey monitor with 21 CFR Part 11 features (OSLOG) software
b) Magnetic door lock feature
c) GSM system: alarms sent to predefined mobile numbers
d) HMI : 5.7” touch screen grey in lieu of 3.7”
e) Controller with printer interface to connect to EPSON Dot matrix line printer. Date, time and temperature can be printed whereas print interval can be set between 1 to 99 minutes
f) PC interface for above controller with software on CD & hardware or software with 21 CFR part 11 features
g) Imported Micro processor based PID controller in lieu of above digital temperature. Controller with Auto tune facility
h) Digital timer of 999 minutes with auto heater cut-off for timed cycle
i) High temperature safety cut-off with audio/visual alarm
Temperature Mapping
Incubator : Performance qualification with temperature mapping on site.
Performance validation test consist of one cycle at any one-temperature point for 2 hours period. This is done to demonstrate the uniformity of temperature within the chamber is acceptable and that the items within chamber receive a uniform temperature condition.
Data is collected in printed format from up to six different points within the chamber.
Temperature uniformity is considered acceptable if the deviation is less ±1ºC of set temperature.
Note: Every OSWORLD Bacteriological Incubator is validated prior to dispatch at the factory. Those customers who do not require on site mapping may request for FREE copy of factory mapping cycle data sheet.