Stability Chamber / Humidity Control Oven


Model No: JRIC 11

 
     
  » Specification  
 

For conditions as per ICH Guidelines

 
  25°C – 60% RH/40°C – 75%RH/30°C – 65%RH / 25°C – 40% RH / 40°C – 25% RH / 30°C – 35% RH.  
       
  Temperature range: 10.0°C to 60.0°C. Accuracy / Uniformity: ±0.1°C/±1°C  
  Humidity range: 25% to 95% RH. Accuracy / Uniformity ±2%RH/±3%RH  
  Temperature / Humidity display: Digital LED 3½ digit  
  Control: Microprocessor based PID control. With Auto tune CE marked  
  Resolution: 0.1°C/1%RH  
  Temperature sensor: PT – 100  
  RH sensor: Direct capacitance type  
  Heating: ‘U’ shaped Nichrome wire heater in SS sheathing  
  Cooling: CFC free Copeland make compressor utilizing R 134A eco friendly refrigerant, with condenser, motor and relay  
  Steam injection: Boiler with reservoir made of thick stainless steel, with heater, water inlet control, low water level safeguard  
  Air circulation: Flange motor with impeller / blower  
  Construction: Double wall, with insulation, outer door key lockable, inner glass viewing door  
  Chamber illumination: Fluorescent light with door switch  
  Safety features  
   
  : High temperature safety cut off
  : Low water level boiler cut off
  : Electrical overload cutoff relay for compressor
  : Time delay for compressor switch ON
  : Electrical circuit breaker
 
  Feet: Castor wheels. Trays: SS wire mesh heavy duty  
  Electrical:  230V / 15A / 50 Hz  
  Calibration certificate by ERTL – Mumbai with traceability to NPL – New Delhi  
 
We understand customer needs as well as aim at exceeding general expectation
 
Constantly strive to strengthen and sustain quality levels
 
36 years old organization with sound integrity and commitment
 
  >> Read more
Stability Chamber
     
 

Pre Installation Mandatory Requirement  :

 
  (Guarantee void incase of non compliance)  
     
  a) Distilled water OR DM water continuous supply with tap for boiler input of ½” BSP at 3 feet height  
  b) Room temperature around machine preferably at 25°C with air conditioning or a well ventilated room with exhaust fan.   
    However surrounding temperature should not exceed 30°C
 
  c) Stabilized input voltage of 230V AC 20 Amps. Use of servo controlled stabilizer is recommended  
  d) Water drain line with ½” nozzle to be provided at floor level  
     
 

STD Model

GMP Model

       
  • Outer cabinet mild steel powder coated
Outer cabinet SS 304 grade dull matt finish
 
 
  • Inner chamber stainless steel 304 grade
Inner chamber stainless steel mirror finish
     
   
With DQ, IQ, OQ documentation
 
     
 
Capacity
Internal Size
External Size
Shelves
Shpg Wt in kgs
Cu.ft
Liters
W x D x H cm
W x D x H cm
(SS)
 
4 120 45 x 45 x 60 65 x 114 x 130 2 200
8 200 60 x 60 x 60 79 x 121 x 130 2 240
12 324 60 x 60 x 90 79 x 121 x 165 3 250
16 450 60 x 60 x 125 79 x 121 x 199 4 265
21 600 60 x 60 x 155 79 x 121 x 247 5 295
28 800 80 x 80 x 125 99 x 141 x 199 5 325
34 1000 80 x 80 x 155 99 x 141 x 247 5 375
 
     
     
 
Optional  Extras / Accessories
a) PLC based system: PLC with HMI 3.7” touch screen grey monitor with 21 CFR Part 11 features (OSLOG) software
b) Magnetic door lock feature
c) GSM system: alarms sent to predefined mobile numbers
d) HMI: 5.7” touch screen grey in lieu of 3.7”
e) Digital PID controller with printer interface to connect EPSON Dot matrix line printer. Print interval programmable, can print date, time, temperature
f) PC interface for above controller with software on CD & hardware or software with 21 CFR part 11 features
g) Imported RH sensor European make, Rotronic Switzerland for higher accuracy
h) Standby humidity system (except for 52 and 120 liters)
i) Standby refrigeration system (except for 52 and 120 liters)
j) Data logger: 4 point temp. + 4 point humidity data logger with sensors placed at specific points in chamber with printer interface and memory of 5000 readings per channel
k) PC interface for above controller with software on CD & hardware
 
     
 

Temperature Mapping

 
     
 

Stability Chamber : Performance qualification with temperature and humidity mapping on site.

Temperature mapping consist of one cycle at any one temperature and humidity point for a 24 hours period. This is done to demonstrate the uniformity of temperature and humidity within the chamber is acceptable and that the samples within the chamber receive a uniform temperature and humidity condition.

Data is collected in printed format from up to eight different points (i.e. six temperature points and two humidity points) within the chamber. Temperature uniformity is considered acceptable if the deviation is less ±1ºC of set temperature. Humidity uniformity is considered acceptable if the deviation is within ± 3% of set humidity.


Note: Every OSWORLD Stability Chamber is validated prior to dispatch at the factory. Those customers who do not require on site mapping may request
for FREE copy of factory mapping cycle data sheet.